Safety and Accuracy
This section refers to the use of Stradwin for 3D ultrasound.
Stradwin is a powerful tool. It is based on the Stradx system that
was first released in 1997 and was shown in 2003 to be the highest
definition freehand three-dimensional ultrasound system in the
literature. Like all powerful tools, its safety and performance is
dependent on correct usage. This page lists some of the issues
you need to be aware of in order to ensure that the software
produces results of sufficient accuracy for your application and
is used as part of a safe system.
Safety
Here is a list of some of the safety issues that you
should consider before using Stradwin.
- The computer that you run Stradwin on must be safe
and must not interfere inappropriately with any other
equipment in the environment.
- Any position sensing device you use must be safe
and must not interfere inappropriately with any other
equipment in the environment.
- The ultrasound machine that you use must be safe
and must not interfere inappropriately with any other
equipment in the environment. Although ultrasound is
generally considered safer than some other imaging modalities,
it does involve risks. Before using any medical imaging
equipment you should be appropriately qualified so that
you do so in an ethical and safe way in the light of the
risks involved.
- Any electrical connections between the equipment you
use should be made by an appropriately qualified person who is
able to ensure they are safe.
- It is dangerous to try and deduce information from
ultrasound images unless you have appropriate qualifications
and experience. Furthermore, the interpretation of some of the
derived images produced by Stradwin requires particular
specialist knowledge over and above the knowledge required for
the interpretation of conventional B-scan images. For example,
ultrasound artifacts appear in a different form in some derived
images.
Configuration
Here is a list of some of the things that need to be considered
when setting up Stradwin.
- The fixed part of the position sensor should be permanently
and rigidly mounted so it cannot be moved either deliberately or
by accident.
- Isocentre calibration should be
performed as described in the documentation.
- Pointer calibration should be performed
as described in the documentation.
- Position sensor mounts should be rigidly attached to all the
probes that are to be used with the system. These mounts must be
rigid and stable. They must be tamper-proof to avoid users moving
them. You will also wish to ensure that they do not void any
warranties or maintenance contracts associated with the ultrasound
machine.
- Make sure that the mobile part of the position sensor is attached
correctly to the mount on the probe. If the users are to be allowed
to move the sensor onto different probes, they need to be taught
to ensure that the sensor is mounted with the correct orientation
(i.e. the same orientation as when the probe calibration was performed).
If the sensor is mounted wrongly the calibration will be invalid.
- All the probes to be used should be
calibrated at all the required depth
settings as described in the documentation. The resulting template
files should be saved somewhere accessible to the intended users.
- All probe calibrations should be tested for consistency with
the pointer calibration, as described in the documentation for the
'Probe Calib' task page.
- Users should be carefully trained never to invert the ultrasound
image, flip it left-to-right or touch the pan and zoom controls. They should always load the
appropriate template file for the probe and depth setting in use.
- Phantom experiments should be performed to check the performance
of the complete system.
- Regularly check that the calibrations of the isocentre,
pointer and all the probes at the various depth settings are still
valid.
User issues
Here is a list of some of the things that need to be considered when
using Stradwin.
- Do not use the system unless it has been properly set up and
configured by an appropriately qualified person.
- Do not use the system unless you have been appropriately
trained.
- If the patient moves during the scan or between the scan and the
radiotherapy treatment, this will seriously compromise the
usefulness of the data. One way to minimise breathing artifacts is
not to mention breathing to the patient and let them just breath
gently and normally. If a patient is trying consciously to breathe in
a particular way, this tends to accentuate rather than reduce the
motion artifacts.
- Always load the correct template corresponding to the probe and
depth setting you are using. If a template does not exist for a
particular combination of probe and depth setting, you should not use
this combination for 3D scans until a template has been prepared.
- Never adjust the way the position sensor is mounted on the probe.
If the mounting is changed, the probe calibration will be invalid.
- Never change the depth setting without loading the appropriate
template file.
- Never use the pan or zoom controls on the ultrasound machine.
- Never flip the B-scan image up-down or left-right.
- Periodically check the consistency of the probe calibrations with
the pointer, as described in the documentation for the
'Probe Calib'
task page. If the calibrations are inconsistent do not use the
system.
Accuracy
Here is a list of some of the sources of error you should consider
when assessing the accuracy of any measurements made using Stradwin.
- Errors caused by the movement of the scanned material.
- Differences between the speed of ultrasound in the object scanned
and the speed assumed by Stradwin and the ultrasound machine you
use.
- The size of the resolution cell (or spatial sensitivity function)
of the overall imaging system.
- Distortion of the tissue due to pressure from the probe during
the scan.
- Errors generated by the algorithms in Stradwin which attempt to
correct for probe pressure distortion.
- Reverberation, shadowing and other ultrasound artifacts.
- The inherent accuracy of the position sensor you use.
- Local disturbances in the performance of the position sensor, for
example optical or magnetic disturbances.
- Distortion introduced in the B-scan when it is transferred as an
analogue video signal.
- Errors in the time-stamping of the image and position
streams.
- Temporal calibration errors - errors in the constants that
determine the relative latencies of the position sensor and
image data streams.
- Spatial calibration of the transformation between the mobile
part of the position sensor and the location of the B-scan in
three-dimensional space.
- Interpolation errors.
- Segmentation errors.
- Errors in the way landmarks are defined either on the data or
using the pointer.
- Interpretive errors.
- Errors caused by delays, persistence or image averaging inside
the ultrasound scanner.
- Errors caused by the approximate temperature
compensation used in the calibration of high curvature probes.